42-229

Rotura de vástago femoral de prótesis total de cadera enteramente recubierta de hidroxiapatita. Presentación de tres casos

Femoral component failure of a total hip arthroplasty with fully coated hidroxiapatite. Presentation of three cases

D. BONETE LLUCH, MC. PULIDO HIGUERAS, C. MAROÑAS ABUELO, JM. GUINOT TORMO, J. PARDO MONTANER SERVICIO DE CIRUGÍA ORTOPÉDICA Y TRAUMATOLOGÍA. HOSPITAL UNIVERSITARIO LA FE DE VALENCIA.
Recepción:
12/09/2008
Aceptación:
12/09/2008
Publicación:
12/09/2008


Métricas


66

Visualizaciones


1

Descargas


Compartir

Compartir

Resumen (ES)

We are describing 3 cases of femoral component
failure of a total hip arthroplasty with fully coated hidroxiapatite
(Furlong, JRI limited, London), without any traumatic
event. The follow-up before failure was 7 and 5 years. The
fracture of femoral stem was always just distal to the metaphisal
cone. The ingrowth of the distal portion of the femoral
component increased by hidrixiapatite causes abnormal
transmission of biomechanical forces in the metafiso-diafisal
union increasing the risk of material failure.

Palabras clave (ES):

Prótesis
Cadera
Complicaciones
Furlong
Rotura de vástago
Artroplastia total

Resumen (EN)

We are describing 3 cases of femoral component
failure of a total hip arthroplasty with fully coated hidroxiapatite
(Furlong, JRI limited, London), without any traumatic
event. The follow-up before failure was 7 and 5 years. The
fracture of femoral stem was always just distal to the metaphisal
cone. The ingrowth of the distal portion of the femoral
component increased by hidrixiapatite causes abnormal
transmission of biomechanical forces in the metafiso-diafisal
union increasing the risk of material failure.

Explorar

Artículos


Volúmenes


Sobre la revista


Más leídos

Surgiceloma como complicación de tenodesis extraarticular de Lemaire modificada asociada a ligamentoplastia del cruzado anterior

Revista Española de Cirugía Osteoarticular


Uso de placa puente dorsal: una alternativa en fracturas de muñeca catastróficas

Revista Española de Cirugía Osteoarticular


No todo es "cut-out": reclasificación de las complicaciones mecánicas del tornillo cefálico del clavo intramedular.

Revista Española de Cirugía Osteoarticular


Solución límite para recambio de prótesis total de cadera con cotilo osteointegrado: presentación de casos

Revista Española de Cirugía Osteoarticular


Radioterapia a bajas dosis en artrosis: ¿una opción terapéutica real o un efecto placebo sofisticado?

Revista Española de Cirugía Osteoarticular