Metalosis en prótesis total de cadera modelo PM
Metallosis after total hip replacement using pm model
V. VICENT CARSI*, M.a J. SANGÜESA NEBOT*, J. J. TERRADEZ RARO** y C. VALVERDE MORDT* * Servicio de Cirugía Ortopédica y Traumatología. ** Servicio de Anatomía Patológica. Hospital Arnau de Vilanova. Valencia.
Recepción:
07/11/2008
Aceptación:
07/11/2008
Publicación:
07/11/2008
Métricas
61
Visualizaciones
Compartir
Compartir
Resumen (ES)
We report a case of a patient who had aseptic loosening of a total hip prosthesis model PM. At revision we found severe metallosis. A black surgery, seven years after the device was implanted pigment that penetrated the capsule and sorrounded the femoral and acetabular components was found. Histological examination of a the specimen of capsular tissue showed histiocytes loaded with small size metallic particles. Of particular interest was the absence of severe clinical and radiological changes. These finds are unusually associated with aseptic loosening níade by polymeric wear debris. The reason of metallosis was supposed tobe the friction between the neck of femoral component and lower rim of the acetabular component.
Palabras clave (ES):
PM
Cadera
Total
Prótesis
Metalosis
Metal
Artroplastia
Resumen (EN)
We report a case of a patient who had aseptic loosening of a total hip prosthesis model PM. At revision we found severe metallosis. A black surgery, seven years after the device was implanted pigment that penetrated the capsule and sorrounded the femoral and acetabular components was found. Histological examination of a the specimen of capsular tissue showed histiocytes loaded with small size metallic particles. Of particular interest was the absence of severe clinical and radiological changes. These finds are unusually associated with aseptic loosening níade by polymeric wear debris. The reason of metallosis was supposed tobe the friction between the neck of femoral component and lower rim of the acetabular component.
Explorar
Artículos
Volúmenes
Sobre la revista
Más leídos
Surgiceloma como complicación de tenodesis extraarticular de Lemaire modificada asociada a ligamentoplastia del cruzado anterior
Revista Española de Cirugía Osteoarticular
Uso de placa puente dorsal: una alternativa en fracturas de muñeca catastróficas
Revista Española de Cirugía Osteoarticular
No todo es "cut-out": reclasificación de las complicaciones mecánicas del tornillo cefálico del clavo intramedular.
Revista Española de Cirugía Osteoarticular
Solución límite para recambio de prótesis total de cadera con cotilo osteointegrado: presentación de casos
Revista Española de Cirugía Osteoarticular
Radioterapia a bajas dosis en artrosis: ¿una opción terapéutica real o un efecto placebo sofisticado?
Revista Española de Cirugía Osteoarticular