Estudio experimental comparativo de la estabilidad de distintos tutores externos
COMPARATIVE EXPERIMENTAL STUDY ON THE STABILITY OF DIFFERENT EXTERNAL
G. VINDVER*, R. AMOR CURTI*, D. PRIETO** y H. CAVIGLIA Departamento de Investigación - División Ortopedia y Traumatología. Hospital General de Agudos «Dr. Juan Fernández».
Recepción:
05/11/2008
Aceptación:
05/11/2008
Publicación:
05/11/2008
Métricas
62
Visualizaciones
Compartir
Compartir
Resumen (ES)
The stiffness of 11 different device configurations of external fixation was investigated in an experimental model using human cadaveric tibia. After application of the different devices, a diaphyseal osteotomy allowing removal of a 35 mm bone segment was performed in the specimens. Displacements of the bone fragments at the osteotomy site induced by compresion loading, torsion, and both anteroposterior and mediolateral flexion were measured with dispositives designed for this experiment. Maximal load and that required for inducing 1 mm displacement were recorded. The higher stiffness corresponded to the Orthofix fixator in axial compression, the Kronner biplanar device in anteroposterior flexion, and the triangular configuration of the AO device in mediolateral flexion. A total stiffness index of each configuration as an average of the behavior against all loading modahties was obtained. The indications, advantages, complications and the biology of bone consolidation wien regards to external fixation are considered. As conclusion, progressive loading of the extremity with unstable fracture requires a close monitorization, since most of the device configurations analyzed allow more than 1 mm displacement of bone fragments at the fracture site, even with low axial loading.
Palabras clave (ES):
Estabilidad.
Tutor.
Fijación.
Externa.
Osteotaxis.
Fijador.
Dinámico.
Experimental.
Resumen (EN)
The stiffness of 11 different device configurations of external fixation was investigated in an experimental model using human cadaveric tibia. After application of the different devices, a diaphyseal osteotomy allowing removal of a 35 mm bone segment was performed in the specimens. Displacements of the bone fragments at the osteotomy site induced by compresion loading, torsion, and both anteroposterior and mediolateral flexion were measured with dispositives designed for this experiment. Maximal load and that required for inducing 1 mm displacement were recorded. The higher stiffness corresponded to the Orthofix fixator in axial compression, the Kronner biplanar device in anteroposterior flexion, and the triangular configuration of the AO device in mediolateral flexion. A total stiffness index of each configuration as an average of the behavior against all loading modahties was obtained. The indications, advantages, complications and the biology of bone consolidation wien regards to external fixation are considered. As conclusion, progressive loading of the extremity with unstable fracture requires a close monitorization, since most of the device configurations analyzed allow more than 1 mm displacement of bone fragments at the fracture site, even with low axial loading.
Explorar
Artículos
Volúmenes
Sobre la revista
Más leídos
Surgiceloma como complicación de tenodesis extraarticular de Lemaire modificada asociada a ligamentoplastia del cruzado anterior
Revista Española de Cirugía Osteoarticular
Uso de placa puente dorsal: una alternativa en fracturas de muñeca catastróficas
Revista Española de Cirugía Osteoarticular
No todo es "cut-out": reclasificación de las complicaciones mecánicas del tornillo cefálico del clavo intramedular.
Revista Española de Cirugía Osteoarticular
Solución límite para recambio de prótesis total de cadera con cotilo osteointegrado: presentación de casos
Revista Española de Cirugía Osteoarticular
Radioterapia a bajas dosis en artrosis: ¿una opción terapéutica real o un efecto placebo sofisticado?
Revista Española de Cirugía Osteoarticular