52-271

Diseño y posicionado de la prótesis inversa de hombro utilizando técnicas de simulación numérica por método de los elementos finitos y biomodelos realizados por impresión 3D

Design and positioning of the reverse shoulder prosthesis using numerical simulation techniques by finite element method and biomodels made by 3D printing

J. Salvador Marín, L. Piles Cabo, J. M. Segui Ripoll, S. Sánchez Caballero, M. J. Reig Pérez, R. Pla Ferrando, S. Ferrándiz Bou, V. J. Seguí Llinares, J. F. Martínez López
Recepción:
09/03/2018
Aceptación:
09/03/2018
Publicación:
09/03/2018


Métricas


76

Visualizaciones


1

Descargas


Compartir

Compartir

Resumen (ES)

To analyze the positioning and orientation of a reverse shoulder prostheses in the scapula, and to improve its anchorage and adaptation using numerical simulation techniques trough finite element methodology (FEM) and biomodels by 3D printing. Methods. A 3D-virtual biomodel is generated by the CT of a male patient of 70 kg and load in a 4,2 kg arm, operated with RSA. The base plate-scapula is studied by FEM software simulating the mechanical behavior, comparing the intervention against eccentricity conditions of 4 mm and lateralization of 5 mm. The tensions and deformations performed on the interface are analyzed. Results. A reduction of the mechanical stress in the base plate of 23.7% is obtained. A guide-centering device, designed using the 3D printing modeling technique, is designed. Conclusion. The FEM modeling and analysis of the patient’s personalized shoulder morphology and 3D biomodel help optimum preoperative planning. Printing 3D guides can improve the positioning and anchoring of the RSA base plate.

Palabras clave (ES):

Estudio
Biodinámico
Planificación
operatoria
Hombro
Elementos
Modelo
Inversa
Prótesis
Finitos
Impresión
3D

Resumen (EN)

To analyze the positioning and orientation of a reverse shoulder prostheses in the scapula, and to improve its anchorage and adaptation using numerical simulation techniques trough finite element methodology (FEM) and biomodels by 3D printing. Methods. A 3D-virtual biomodel is generated by the CT of a male patient of 70 kg and load in a 4,2 kg arm, operated with RSA. The base plate-scapula is studied by FEM software simulating the mechanical behavior, comparing the intervention against eccentricity conditions of 4 mm and lateralization of 5 mm. The tensions and deformations performed on the interface are analyzed. Results. A reduction of the mechanical stress in the base plate of 23.7% is obtained. A guide-centering device, designed using the 3D printing modeling technique, is designed. Conclusion. The FEM modeling and analysis of the patient’s personalized shoulder morphology and 3D biomodel help optimum preoperative planning. Printing 3D guides can improve the positioning and anchoring of the RSA base plate.

Explorar

Artículos


Volúmenes


Sobre la revista


Más leídos

Surgiceloma como complicación de tenodesis extraarticular de Lemaire modificada asociada a ligamentoplastia del cruzado anterior

Revista Española de Cirugía Osteoarticular


Uso de placa puente dorsal: una alternativa en fracturas de muñeca catastróficas

Revista Española de Cirugía Osteoarticular


No todo es "cut-out": reclasificación de las complicaciones mecánicas del tornillo cefálico del clavo intramedular.

Revista Española de Cirugía Osteoarticular


Solución límite para recambio de prótesis total de cadera con cotilo osteointegrado: presentación de casos

Revista Española de Cirugía Osteoarticular


Radioterapia a bajas dosis en artrosis: ¿una opción terapéutica real o un efecto placebo sofisticado?

Revista Española de Cirugía Osteoarticular