Fallo recurrente tipo IV de un inserto acetabular retentivo sin disociación del anillo de bloqueo.

A. LIZAUR UTRILLA, D. BUSTAMANTE SUÁREZ DE PUGA
Recepción:
22/06/2011
Aceptación:
22/06/2011
Publicación:
23/06/2011


Métricas


58

Visualizaciones


Compartir

Compartir

Resumen (ES)

We report recurrent dislocation with two failed Duofit IPR constrained liners in the same hip. Both times, the cementless cup and stem were well fixed and well positioned, and the constrained liner was found to be well fixed within the cup and undamaged. The locking ring was well positioned with complete seating into the liner, but had a small "scratch" and a plastic deformation of superior and inferior poles, respectively. This finding suggested that impingement between the neck and the rim of the ring had occurred, especially during flexion, so-called failure type IV. It required revision to a large-diameter metal-on-metal modular articulation. The well-fixed and wellpositioned femoral stem was left in place. We are not aware of any other published reports concerning failure of this implant.

Palabras clave (ES):

Artroplastia
Cadera
Acetábulo
retentivo
Luxación
Dislocación
Prótesis

Resumen (EN)

We report recurrent dislocation with two failed Duofit IPR constrained liners in the same hip. Both times, the cementless cup and stem were well fixed and well positioned, and the constrained liner was found to be well fixed within the cup and undamaged. The locking ring was well positioned with complete seating into the liner, but had a small "scratch" and a plastic deformation of superior and inferior poles, respectively. This finding suggested that impingement between the neck and the rim of the ring had occurred, especially during flexion, so-called failure type IV. It required revision to a large-diameter metal-on-metal modular articulation. The well-fixed and wellpositioned femoral stem was left in place. We are not aware of any other published reports concerning failure of this implant.

Explorar

Artículos


Volúmenes


Sobre la revista


Más leídos

Surgiceloma como complicación de tenodesis extraarticular de Lemaire modificada asociada a ligamentoplastia del cruzado anterior

Revista Española de Cirugía Osteoarticular


Uso de placa puente dorsal: una alternativa en fracturas de muñeca catastróficas

Revista Española de Cirugía Osteoarticular


No todo es "cut-out": reclasificación de las complicaciones mecánicas del tornillo cefálico del clavo intramedular.

Revista Española de Cirugía Osteoarticular


Solución límite para recambio de prótesis total de cadera con cotilo osteointegrado: presentación de casos

Revista Española de Cirugía Osteoarticular


Radioterapia a bajas dosis en artrosis: ¿una opción terapéutica real o un efecto placebo sofisticado?

Revista Española de Cirugía Osteoarticular