40-222

Vástago spotorno. Resultados a 10 años

Spotorno stem. Results of a 10 year follow-up study

FJ. MODREGO ARANDA SERVICIO DE TRAUMATOLOGÍA Y CIRUGÍA ORTOPÉDICA. HOSPITAL DE BARBASTRO.
Recepción:
15/09/2008
Aceptación:
15/09/2008
Publicación:
15/09/2008


Métricas


60

Visualizaciones


1

Descargas


Compartir

Compartir

Resumen (ES)

Spotorno stem is a tapered straight cementless
stem; Spotorno stem achieved initial stability by wedging a
proximal fluted, straight stem into a retained bed of femoral
trabecular and cortical bone, rectangular cross section provides
excellent rotational stability distally. The titanium-aluminum-
niobium alloy enables the fast ongrowth of newly formed
osseous tissue to act as secondary stabilization This retrospective
study reviews 26 Spotorno stems, the mean follow-up
was 134 months. Clinical results were graded excellent and
good in 93 per cent of hips. At follow-up, none patient presented
radiological evidence of aseptic stem loosening and 100
per cent of stems showed stable fixation.

Palabras clave (ES):

Artroplastia total
Vástago
Spotorno
No cementado
Resultados
Cadera
Prótesis

Resumen (EN)

Spotorno stem is a tapered straight cementless
stem; Spotorno stem achieved initial stability by wedging a
proximal fluted, straight stem into a retained bed of femoral
trabecular and cortical bone, rectangular cross section provides
excellent rotational stability distally. The titanium-aluminum-
niobium alloy enables the fast ongrowth of newly formed
osseous tissue to act as secondary stabilization This retrospective
study reviews 26 Spotorno stems, the mean follow-up
was 134 months. Clinical results were graded excellent and
good in 93 per cent of hips. At follow-up, none patient presented
radiological evidence of aseptic stem loosening and 100
per cent of stems showed stable fixation.

Explorar

Artículos


Volúmenes


Sobre la revista


Más leídos

Surgiceloma como complicación de tenodesis extraarticular de Lemaire modificada asociada a ligamentoplastia del cruzado anterior

Revista Española de Cirugía Osteoarticular


Uso de placa puente dorsal: una alternativa en fracturas de muñeca catastróficas

Revista Española de Cirugía Osteoarticular


No todo es "cut-out": reclasificación de las complicaciones mecánicas del tornillo cefálico del clavo intramedular.

Revista Española de Cirugía Osteoarticular


Solución límite para recambio de prótesis total de cadera con cotilo osteointegrado: presentación de casos

Revista Española de Cirugía Osteoarticular


Radioterapia a bajas dosis en artrosis: ¿una opción terapéutica real o un efecto placebo sofisticado?

Revista Española de Cirugía Osteoarticular